2
SCPI Device Message Details
[:SENSe]:CHPower:SWEep:TIME:AUTO?
Sweep Time Auto/Manual Query
Function
This command queries the On/Off state of the automatic sweep time
setting.
Refer to
[:SENSe]:SWEep:TIME:AUTO?.
Related command
This command has the same function as the following commands.
[:SENSe]:SWEep:TIME:AUTO?
[:SENSe]:ACPower:SWEep:TIME:AUTO?
[:SENSe]:OBWidth:SWEep:TIME:AUTO?
:CALCulate:CHPower:MARKer[1]|2|3|4|5|6|7|8|9|10:REFerence <integer>
Relative To
Function
This command sets the reference marker when the marker mode is set to
Delta.
Refer to
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:REFerence.
Related command
This command has the same function as the following commands.
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:REFerence
:CALCulate:ACPower:MARKer[1]|2|3|4|5|6|7|8|9|10:REFerenc
e
:CALCulate:OBWidth:MARKer[1]|2|3|4|5|6|7|8|9|10:REFerenc
e
:CALCulate:BPOWer|:TXPower:MARKer[1]|2|3|4|5|6|7|8|9|10:
REFerence