SCPI Device Message Details
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel <rel_ampl>
Reference Level (Normalize)
Function
This command sets the Reference Level when the Normalize finction is
performed.
Command
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel <rel_ampl>
Parameter
<rel_ampl>
Reference Level
Range –120.00 to 50.00 dB
Resolution 0.01 dB
Suffix code DB, dB is used even when omitted
Default 0 dB
Details
This command is available only for MS2830A.
Example of Use
To set Reference Level to 10 dB.
DISP:WIND:TRAC:Y:NRL 10.00DB
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel?
Reference Level (Normalize) Query
Function
This command queries Reference Level when the Normalize function is
performed.
Query
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:NRLevel?
Response
<rel_ampl>
Parameter
<rel_ampl>
Reference Level
Range –120.00 to 50.00 dB
Resolution 0.01 dB
Suffix code None. Value is returned in dB units.
Details
This command is available only for MS2830A.
Example of Use
To query Reference Level.
DISP:WIND:TRAC:Y:NRL?
> 10.00