Chapter
Native Device Message Details
4-36
CAL:RCL/CAL:RCL?
Adjust Reference Clock
Function
This command sets the adjustment value for the internal reference signal
oscillator (Reference Clock).
Command
CAL:RCL integer
Query
CAL:RCL?
Response
integer
Parameter
integer
Adjustment value
Range 0 to 1023 (MS2830A)
0 to 4095 (MS2840A)
Resolution 1
Example of Use
To set the adjustment value of the internal reference signal oscillator to
511
CAL:RCL 511
CAL:RCL?
> 511
CAL:RCL:PRES
Adjust Reference Clock Preset
Function
This command resets the adjustment value of the internal reference
signal oscillator (Reference Clock).
Command
CAL:RCL:PRES
Example of Use
To reset the adjustment value of the internal reference signal oscillator.
CAL:RCL:PRES