Chapter 2 SCPI Device Message Details
2-368
:CALCulate:ACPower:CARRier[1]:LIST:BANDwidth[:INTegration]?
Adjacent Channel Power Carrier Bandwidth Query
Function
This command sets the measurement bandwidth of the carrier for
Adjacent Channel Power measurement.
Refer to
[:SENSe]:ACPower:CARRier[1]:LIST:BANDwidth[:INTegration]
?.
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:CARRier[1]:LIST:BANDwidth[:INTegration]
?
[:SENSe]:ACPower:CARRier[1]:RCFRequency <freq>
Adjacent Channel Power In Band Center
Function
This command sets the In Band center frequency for Adjacent Channel
Power measurement.
Command
[:SENSe]:ACPower:CARRier[1]:RCFRequency <freq>
Parameter
<freq>
In Band center frequency
Range (Start Frequency) to (Stop Frequency)
Resolution 1 Hz
Suffix code HZ,KHZ,KZ,MHZ,MZ,GHZ,GZ
Hz is used when omitted.
Default Center Frequency
Example of Use
To set In Band center frequency to 3 GHz.
ACP:CARR:RCFR 3GHZ
Related command
This command has the same function as the following command.
:CALCulate:ACPower:CARRier[1]:RCFRequency