2
SCPI Device Message Details
[:SENSe]:FREQuency:STARt?
Start Frequency Query
Function
This command queries the start frequency.
Query
[:SENSe]:FREQuency:STARt?
Response
<freq>
Parameter
<freq>
Start frequency
Range
[MS269xA]
–100 MHz to 6.0499997 GHz (MS2690A)
–100 MHz to 13.5999997 GHz (MS2691A)
–100 MHz to 26.5999997 GHz (MS2692A)
[MS2830A]
–100 MHz to 3.6999997 GHz (Option 040)
–100 MHz to 6.0999997 GHz (Option 041)
–100 MHz to 13.5999997 GHz (Option 043)
–100 MHz to 26.5999997 GHz (Option 044)
–100 MHz to 43.0999997 GHz (Option 045)
[MS2840A]
–100 MHz to 3.6999997 GHz (Option 040)
–100 MHz to 6.0999997 GHz (Option 041)
–100 MHz to 26.9999997 GHz (Option 044)
–100 MHz to 44.9999997 GHz (Option 046)
Resolution 1 Hz
Suffix code None. Value is returned in Hz units.
Example of Use
To query the start frequency.
FREQ:STAR?
> 1000000000