Chapter 2 SCPI Device Message Details
2-638
[:SENSe]:SEMask:SWEep[1]:POINts?
Spectrum Emission Mask Trace Point Query
Function
This command queries the number of frequency display points when
measuring the reference power for Spectrum Emission Mask
measurement.
Query
[:SENSe]:SEMask:SWEep[1]:POINts?
Response
<integer>
Parameter
<integer>
Number of frequency display points
Range 11 to 10001
Resolution Any of the following values can be selected:
11, 21, 41, 51, 101, 201, 251, 401, 501, 1001,
2001, 5001, 10001, 30001 (MS269xA Only)
Suffix code None
Example of Use
To query the number of frequency display points when measuring the
reference power.
SEM:SWE:POIN?
> 2001