Chapter 2 SCPI Device Message Details
2-236
:MMEMory:LOAD:LLINe[1]|2|3|4|5|6 <filename>[,<device>]
Recall Limit Data file
Function
This command recalls the contents of the Limit data file.
Command
:MMEMory:LOAD:LLINe[1]|2|3|4|5|6 <filename>[,<pass>]
Parameter
<filename> "
Name of file to recall
"
String of up to 32 characters, not including the
extension, enclosed in double quotes (") or single
quotes (')
The following characters cannot be used:
\ / : * ? “ ” ‘ ’ < > |
<device>
Drive name
A,B,D,E,F,...
Defaults to D when omitted.
Details
This command can recall Limit files in csv, lim, or xml format.
From the following folder in the specified drive, the specified file is
loaded:
\Anritsu Corporation\Signal Analyzer\User Data\Limit
Example of Use
To load the Limit data file “Limit.csv from E drive to Limit3.
MMEM:LOAD:LLIN3 "Limit.csv",E