Chapter 2 SCPI Device Message Details
2-156
:CALCulate:MARKer[1]|2|3|4|5|6|7|8|9|10:FCOunt[:STATe] OFF|ON|0|1
Frequency Counter State
Function
This command turns On/Off the frequency counter.
Command
:CALCulate:MARKer[n]:FCOunt[:STATe] <switch>
Parameter
<n>
Marker number
1
Marker 1
2
Marker 2
3
Marker 3
4
Marker 4
5
Marker 5
6
Marker 6
7
Marker 7
8
Marker 8
9
Marker 9
10
Marker 10
When omitted: Marker 1
<switch>
Frequency counter On/Off
ON|1
Sets marker n as the active marker and switches
on the frequency counter function.
OFF|0
Switches off the frequency counter function of
marker n.
Details
The frequency counter is available only for the active markers. Following
execution of this command, the marker with the specified marker
number becomes active.
The setting value is shared by all the markers.
This command is not available when Gate Sweep is On.
This command is not available when Marker Result is Integration or Density.
This command is not available when RBW is equal to or lower than 30 Hz.
This command is disabled during the Spurious Emission measurement.
This command is disabled during the Spectrum Emission Mask measurement.
This command is disabled when Noise Cancel is set to On.
This command is disabled when Internal Signal Generator Control
On/Off is set to On.
Example of Use
To set marker 1 as the active marker and switch on the frequency
counter function.
CALC:MARK:FCO ON