Chapter 2 SCPI Device Message Details
2-624
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME:AUTO
<switch_1>[,<switch_2>[,<switch_3>[,<switch_4>[,<switch_5>[,<switch_6>[,<
switch_7>[,<switch_8>[,<switch_9>[,<switch_10>[,<switch_11>[,<switch_12>]
]]]]]]]]]]
Spectrum Emission Mask Offset Sweep Time Auto/Manual
Function
This command enables/disables the automatic sweep time setting
function for the offset during Spectrum Emission Mask measurement.
Command
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME:AUTO
<switch_n>
Parameter
<switch_n>
Automatic sweep time setting function On/Off
for Offset-n
ON|1
Enables the automatic sweep time setting
function.
OFF|0
Disables the automatic sweep time setting
function.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To enable the automatic sweep time setting function for the offset.
SEM:OFFS:LIST:SWE:TIME:AUTO
OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF,OFF