Chapter 2 SCPI Device Message Details
2-620
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME
<time_1>[,<time_2>[,<time_3>[,<time_4>[,<time_5>[,<time_6>[,<time_7>[,<ti
me_8>[,<time_9>[,<time_10>[,<time_11>[,<time_12>]]]]]]]]]]]
Spectrum Emission Mask Offset Sweep Time
Function
This command sets the sweep time of the offset for Spectrum Emission
Mask measurement.
Command
[:SENSe]:SEMask:OFFSet[1]:LIST:SWEep[1]:TIME <time_n>
Parameter
<time_n>
Sweep time of Offset-n
Range
[MS269xA]
2 ms to 1000 s
[MS2830A]
1 ms to 1000 s
[MS2840A]
1 ms to 1000 s
Resolution 1 ms (time ≤ 1 s)
0.1 s (1 s < time < 10 s)
1 s (10 s ≤ time)
Suffix code NS,US,MS,S
S is used when omitted.
Details
The setting range of this function is limited depending on the settings of
other parameters. Refer to the
MS2690A/MS2691A/MS2692A Signal
Analyzer Operation Manual (Spectrum Analyzer Function Operation)
or
MS2830A/MS2840A Signal Analyzer Operation Manual (Spectrum
Analyzer Function Operation)
for details.
Example of Use
To set the sweep time of the offset.
SEM:OFFS:LIST:SWE:TIME
40MS,40MS,40MS,40MS,40MS,40MS,40MS,40MS,40MS,40MS,40MS,4
0MS