2
SCPI Device Message Details
:CALCulate:MARKer:PEAK:RESolution|EXCursion <rel_ampl>|<percent>
Marker Peak Excursion
Function
This command sets the resolution in detecting the peak point.
Command
When Scale Mode is set to Log.
:CALCulate:MARKer:PEAK:RESolution|EXCursion <rel_ampl>
When Scale Mode is set to Lin.
:CALCulate:MARKer:PEAK:RESolution|EXCursion <percent>
Parameter
<rel_ampl>
Range 0.001 to 100.000 dB
Suffix code DB
dB is used even when omitted.
Default 2 dB
<percent>
Range 0.01 to 100.00%
Suffix code None
Default 2%
Details
This command is not available during Spurious Emission measurement
and when Displayed Segment Mode is Auto.
Not available during Spectrum Emission Mask measurement.
Example of Use
To set the resolution to 20 dB.
CALC:MARK:PEAK:RES 20DB