Chapter 2 SCPI Device Message Details
2-36
Example of Use
To query the center frequency.
FREQ:CENT?
> 123456
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
Frequency Step Size
Function
This command sets the step size of the center, start and stop frequency.
Command
[:SENSe]:FREQuency:CENTer:STEP[:INCRement] <freq>
Parameter
<freq>
Step size
Range
[MS269xA]
1 Hz to 6.00 GHz (MS2690A)
1 Hz to 13.5 GHz (MS2691A)
1 Hz to 26.5 GHz (MS2692A)
[MS2830A]
1 Hz to 3.6 GHz (Option 040)
1 Hz to 6.0 GHz (Option 041)
1 Hz to 13.5 GHz (Option 043)
1 Hz to 26.5 GHz (Option 044)
1 Hz to 43 GHz (Option 045)
[MS2840A]
1 Hz to 3.6 GHz (Option 040)
1 Hz to 6.0 GHz (Option 041)
1 Hz to 26.5 GHz (Option 044)
1 Hz to 44.5 GHz (Option 046)
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
Default value 1 GHz
Example of Use
To set the step size to 100.0 kHz.
FREQ:CENT:STEP 100000