2.9 ACP
2-375
2
SCPI Device Message Details
[:SENSe]:ACPower:OFFSet[1]:LIST:STATe
ON|OFF|1|0,ON|OFF|1|0,ON|OFF|1|0
Adjacent Channel Power Offset
Function
This command sets offset channel On/Off for ACP measurement.
Command
[:SENSe]:ACPower:OFFSet[1]:LIST:STATe
<switch_1>,<switch_2>,<switch_3>
Parameter
<switch_n>
Measurement channel “n” On/Off
ON|1
Measures offset channel “n”.
OFF|0
Does not measure offset channel “n”.
Default
offset-1
On
offset-2
On
offset-3
Off
Example of Use
To set offset channel.
ACP:OFFS:LIST:STAT ON,ON,ON
Related command
This command has the same function as the following command.
:CALCulate:ACPower:OFFSet[1]:LIST:STATe
:CALCulate:ACPower:OFFSet[1]:LIST:STATe
ON|OFF|1|0,ON|OFF|1|0,ON|OFF|1|0
Adjacent Channel Power Offset
Function
This command sets offset channel On/Off for Adjacent Channel Power
measurement.
Refer to
[:SENSe]:ACPower:OFFSet[1]:LIST:STATe.
Related command
This command has the same function as the following command.
[:SENSe]:ACPower:OFFSet[1]:LIST:STATe