EasyManua.ls Logo

Anritsu MS2830A - :TRIGger[:SEQuence]:HOLDoff <time>

Anritsu MS2830A
1452 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 2 SCPI Device Message Details
2-284
:TRIGger[:SEQuence]:HOLDoff <time>
Trigger Hold
Function
This command sets the time during which trigger input is disabled for a
set time, from the first trigger input until the next trigger input.
Command
:TRIGger[:SEQuence]:HOLDoff <time>
Parameter
<time> Offset time
Range 0 to 1 s
Resolution 10 ns
Suffix code NS, US, MS, S
S is used when the suffix code is omitted.
Default 100 µs
Details
This command is not available for MS269x Series.
The Trigger Hold (ON/OFF) function is set to ON when the value is
changed with this function.
This function cannot be set in the case of video trigger.
Example of Use
To set the trigger input disable time to 100 ms.
TRIG:HOLD 100MS

Table of Contents

Other manuals for Anritsu MS2830A

Related product manuals