Chapter
4
Native Device Message Details
4-29
4
Native Device Message Details
BAND:MODE/BAND:MODE?
Resolution Bandwidth Normal/CISPR
Function
This command switches the Resolution Bandwidth mode. When Couple
Time/Freq. Domain is set to Off, each parameter is specified in
accordance with the already-specified frequency axis or time axis
measurement.
Command
BAND:MODE mode
Query
BAND:MODE?
Response
mode
Parameter
mode
NORM
RBW (in normal cases). This is the default value.
CISP
CISPR RBW
Details
This function is available only when Option 016/116 is installed for
MS2830A, MS2840A.
Not available when the Measure function is set to On.
Not available when the Gate View function is set to On.
Example of Use
To switch the mode to CISPR RBW.
BAND:MODE CISP
BAND:MODE?
> CISP