2.6 Trace
2-235
2
SCPI Device Message Details
:MMEMory:STORe:LLINe[1]|2|3|4|5|6 [<filename>[,<device>]]
Save Limit
Function
This command saves the Limit data settings to a CSV file.
Command
:MMEMory:STORe:LLINe[1]|2|3|4|5|6
[,<filename>[,<device>]]
Parameter
<filename> "
Name of file to save
"
String of up to 32 characters, not including the
extension, enclosed in double quotes (") or single
quotes (')
The following characters cannot be used:
\ / : * ? “ ” ‘ ’ < > |
When omitted it is saved as a csv file with the
following default file name: Limit
“Date”_“Sequential Number”.
<device>
Drive name
A,B,D,E,F,...
Defaults to D when omitted.
Details
When the file name is omitted, a sequential number from 0 to 999 is
added to the default file name. When the sequential number reaches
999, no more files can be saved.
Files are saved to the following directory in the specified drive.
\Anritsu Corporation\Signal Analyzer\User Data\Limit
Up to 1000 files can be stored in the folder.
Example of Use
To save the Limit data "Limit" of Limit3 to E drive.
MMEM:STOR:LLIN3 "Limit",E