Chapter 2 SCPI Device Message Detail
2-960
:CALibration:RCLock[:VALue]:PRESet
Adjust Reference Clock Preset
Function
This command resets the adjustment value of the internal reference
signal oscillator (Reference Clock).
Command
:CALibration:RCLock[:VALue]:PRESet
Example of Use
To reset the adjustment value of the internal reference signal oscillator.
CAL:RCL:PRES
[:SENSe]:POWer[:RF]:MW:PRESelector[:STATe] ON|OFF|1|0
Micro Wave Preselector Bypass
Function
This command sets the Micro Wave Preselector Bypass.
Command
[:SENSe]:POWer[:RF]:MW:PRESelector[:STATe] <switch>
Parameter
<switch>
Microwave Preselector Bypass
ON|1
Enables bypassing
OFF|0
Disables bypassing
Default
OFF
Details
This function is available when Option 067/167 is installed for MS269xA.
This function is available when Option 007/067/167 is installed for
MS2830A.
This function is available when Option 067/167 is installed for MS2840A.
Example of Use
To set the Micro Wave Preselector Bypass.
POW:MW:PRES ON