2
SCPI Device Message Details
:CALCulate:LLINe[1]|2|3|4|5|6:DATA
<x-axis_1>,<ampl_1>,<connected_1>[,<x-axis_2>,<ampl_2>,<connected_2>]
,,,[,<x-axis_n>,<ampl_n>,<connected_n>]
Limit Edit
Function
This command sets the
,
, and
parameters for Limit Points on the specified Limit Line.
Command
:CALCulate:LLINe[1]|2|3|4|5|6:DATA
<x-axis_1>,<ampl_1>,<connected_1>[,<x-axis_2>,<ampl_2>,<
connected_2>],,,[,<x-axis_n>,<ampl_n>,<connected_n>]
Parameter
<x-axis_n>
Frequency Offset
Range 0 to 325 GHz (Fixed/Absolute)
–100 GHz to 100 GHz (Relative)
Default Start Frequency (Fixed/Absolute)
0 Hz (Relative)
Resolution 1 Hz
Suffix code HZ, KHZ, KZ, MHZ, MZ, GHZ, GZ
Hz is used when omitted.
<
ampl
_n>
Amplitude
Range
–300 to dBm (Fixed/Absolute)
–300 to 300 dB (Relative)
Default Reference Level (Fixed/Absolute)
0 dB (Relative)
Resolution 0.01 dB
Suffix code None
<connected_n>
Connected to Previous point
OFF|0
Disabled (No Connected)
ON|1
Enabled (Connected) (Default)
Details
Up to 100 points can be set.
When you try to set more than 100 points, the error message
“Insufficient data” is displayed.
When a Limit point’s Frequency value is higher than the other Limit
point’s value and this relationship reverses, the order of Limit points
changes.
Example of Use
To set on Limit line1 to 1 GHz, –20 dBm, and No Connected for Point1,
and set to 2 GHz, –30 dBm, and Connected for Point2.
CALC:LLIN1:DATA 1000000000,-20,0,2000000000,-30,1