Chapter 2 SCPI Device Message Details
2-190
:TRACe[1]|2|3|4|5|6:STORage:MODE
OFF|MAXHold|AVERage|MINHold|LAVerage
Storage Mode
Function
This command sets the storage mode of trace data.
Command
:TRACe[n]:STORage:MODE <mode>
Parameter
<n>
Target trace
1
Trace A
2
Trace B
3
Trace C
4
Trace D
5
Trace E
6
Trace F
When omitted: Trace A
<mode>
Storage mode
OFF
Does not store trace data.
MAXHold
Holds the maximum value.
AVERage
Executes Log averaging.
MINHold
Holds the minimum value.
LAVerage
Executes Linear averaging.
Details
Trace B, C, D, E, and F cannot be set during the Spectrum Emission
Mask measurement.
Trace B, C, D, E, and F cannot be set during the Spurious Emission
measurement.
Example of Use
To set the storage mode of Trace B to Average.
TRAC2:STOR:MODE AVER
Related command
This command has the same function as the following commands.
[:SENSe]:ACPower:AVERage[:STATe]
[:SENSe]:CHPower:AVERage[:STATe]
[:SENSe]:OBWidth:AVERage[:STATe]
[:SENSe]:SEMask:AVERage[:STATe]
[:SENSe]:BPOWer|:TXPower:AVERage[:STATe]
[:SENSe]:SPURious:AVERage[:STATe]