Native Device Message Details
4-48
time_n
Time of marker n
No suffix code, in s units, resolution: 0.1 ns
Returns 999999999999 when no measurement is performed, an
error occurs, or the marker is off.
Details
This command is not available during the Spurious Emission
measurement and when Displayed Segment Mode is set to Auto.
Example of Use
To query all marker values (frequency domain)
CALC:MARK:READ?
>
1000000.00,-15.321,1100000.00,-23.000,1200000.00,-15.321
,1300000.00,-12.680,1400000.00,-5.622,1500000.00,-65.056
,1600000.00,-26.534,1700000.00,-34.264,1800000.00,-35.64
4,-999999999999,-999.0