Chapter 2 SCPI Device Message Details
2-414
Parameter
<lower_offset_n_rel>
Relative power of the lower Offset-n
<upper_offset_n_rel>
Relative power of the upper Offset-n
No suffix code, dB units, 0.01 dB resolution
–999.0 is returned when an error occurs/no
measurement is performed.
<ref_carrier_b>
Power of the reference carrier
<ref_carrier_lower>
Power of the carrier on the left
<ref_carrier_upper>
Power of the carrier on the right
<total_carrier_b>
Total power of all the carriers
<channel_n_rel>
Relative power of the carrier “n” to the
carrier “m”.
However, “m” is a value rounded out
from Carrier Number
÷
2.
<lower_offset_n_abs>
Absolute power of the lower Offset-n
<upper_offset_n_abs>
Absolute power of the upper Offset-n
No suffix code, dBm unit, 0.01 dB resolution
–999.0 is returned when an error occurs/no
measurement is performed.
<integration_abs>
Absolute value of the integral power on
the whole screen
<ref_carrier_a>
Power of the reference carrier
<total_carrier_a>
Total power of all the carriers
<power_n>
Absolute value of the power of carrier
–n
Without a suffix code, Log Scale Unit
(Note: dBm units for V, W units for W)
−
999.0 is returned when an error has occurred or
no measurement is performed.
Details
This function outputs the measurement result of the ACP measurement
last performed. It is possible to output the measurement result in a state
that the measurement has already been done, and in a different style.
You can use READ command if you perform a remeasurement along with
starting a sweep again.
The return values of this function change according to the result mode.
(cf.
:SYSTem:RESult:MODE
)
Example of Use
To query the measurement result of ACP measurement.
FETC:ACP?
>
0.0,-72.130,0.0,-72.130,-1.270,-73.400,-0.570,-72.700,-0
.780,-72.910,-1.030,-73.160,-999.0,-999.0,-999.0,-999.0