Chapter 2 SCPI Device Message Detail
2-798
<number>
(n=22)
<tracedata_1>,<tracedata_2>
・・・・
<tracedata_m>
(n=23 to 42)
Parameter
<spur_n>
Spurious number
<range_n>
Segment number of the detected spurious
<number>
Number of the detected spurious
999.0 is returned when an error occurs/no
measurement is performed.
<freq_n>
Frequency of Spurious
Values have no suffix code, are in Hz units, and
have 0.01 Hz Resolution.
Only one value is returned when Marker Result
Type is Worst.
999999999999 is returned when an error
occurs/no measurement is performed.
<peak_n>
Absolute power of Spurious
Values have no suffix code, are in dBm units,
and have 0.01 dB Resolution.
Only one value is returned when Marker Result
Type is Worst.
999.0 is returned when an error occurs/no
measurement is performed.
<margin_n>
Relative power from the limit line of Spurious
Values have no suffix code, are in dB units, and
have 0.01 dB Resolution.
Only one value is returned when Marker Result
Type is Worst.
999.0 is returned when an error occurs/no
measurement is performed.
<limit_n>
Power value of the limit line of Spurious
Values have no suffix code, are in dBm units,
and have 0.01 dB Resolution.
Only one value is returned when Marker Result
Type is Worst.
999.0 is returned when an error occurs/no
measurement is performed.
<judge_n>
Limit line judges the detected spurious as Pass
or Fail.