2.17 Configuring Batch Measurement Settings
2-947
2
SCPI Device Message Details
<time>
Power adjustment time
Suffix code
None, ms (unit)
–999.0 is returned if there is no measurement.
<sa_input_log_n>
DUT output level at power adjustment
Added to response when Log Output on
Suffix code
None, dBm (unit)
0.01 dB (resolution)
<sg_output_log_n>
Output level from SG Output at each power
adjustment
Added to response when Log Output on
<count_log_n>
Power adjustment count
Added to response when Log Output on
Suffix code
None, dBm (unit)
0.01 dB (resolution)
Details
This function cannot be performed when the SG option is not installed.
SG setting and measurement are executed repeatedly until the target
level specified at target, range is reached within the count specified at
count. This is used for various measurements such as a spurious
measurement with specified output level.
In this application, the Reference Level and Attenuator settings are
calculated as follows from the Adjustment Target Level, Crest Factor,
and Correction value and are set automatically at the measurement
section. Using this application, the Attenuator is set so that the Mixer
Input Level become 0 dBm.
Reference Level setting value
Correction -
Factor Crest target Level Reference +><=
* Crest Factor = 12 dB
However, it is rounded to 50 dBm when Reference Level > 50 dBm (30
dBm at Pre-amp On) and to –120 dBm when Reference Level < –120
dBm
Attenuator setting values
Offset-Gain Amp-
Pre LevelInput
Mixer
- Level Reference
+=Attenuator
* Mixer Input Level = 0
* Pre-Amp Gain = 20 dB (only at Pre-Amp On)
* Offset :Reference Level Offset value
However, it is rounded to 60 dB when Attenuator > 60 and to 0 dBm
when Attenuator < 0. Additionally, when the Attenuator value is an odd