When this function is
enabled
, Network Master will reply to incoming ARP
requests, using the Source MAC and IP addresses in sending out replies.
ARP: Address Resolution Protocol
Note that when using the Ping application, this setting is ignored, and replies
will always be sent - also from the passive port.
Accept Network Master configuration framesAccept Network Master configuration frames
This setting appears in the RFC2544 and SAT1564 applications.
When the check box is selected, Network Master can be configured by another
Network Master via test interface.
Receiver SetupReceiver Setup
Expected preamble lengthExpected preamble length
Allows you to specify the preamble length that the receiver should consider as
'normal' that is, non-erroneous). The default setting is 88 bytes. The range is
from 33 to 1515 bytes.
Ignore preamble violationsIgnore preamble violations
Select this check box if you want to ignore preamble violations.
IFG lower thresholdIFG lower threshold
Allows you to specify the minimum allowed receiver interframe gap. The
default setting is 1212 bytes. The range is from 88 to 1515 bytes.
Filter IFG violations caused by master/slave clock synchronizationFilter IFG violations caused by master/slave clock synchronization
Applies only to Gigabit mode, where the IFG will not be constant, which leads
to a larger number of IFG violated frames than expected. Select this function to
hide those violations.
Jumbo frame size upper limitJumbo frame size upper limit
Allows you to specify the size of the Jumbo frames (that is, Ethernet frames
longer than 1518 that are not counted as oversized/erroneous). The default
setting is 90189018 bytes. The range is from 15191519 to 1600016000 bytes.
In-band SetupIn-band Setup
This tab provides the settings for the remote control using commands via the
test interface.
In-band control communication supports the following protocols.
Ethernet/IPv4/TCP
Ethernet/VLAN/IPv4/TCP
The TCP port number is set in Remote Control on the Instrument Tool bar.
Allow In-band controlAllow In-band control
Selecting the check box enables the remote control using commands via test
interface.