General Specifications
Source Operation
Worst-case accuracy specifications simultaneously include all errors
contributed by variations in power line voltage, load regulation, stability,
and temperature, up to full output power. Includes stable source
operation in four quadrants while delivering power — load power factor
from nearly 1 to 0, leading or lagging. Each F6000 Instrument is supplied
with a Certificate of Calibration traceable to the National Institute of
Standards and Technology.
Electrostatic Discharge Immunity
IEC 801-2 I.E.C. performance level 1 @ 10 kV: normal performance
within specifications. IEC 801-2 I.E.C. performance level 2 @ 20 kV:
no permanent damage.
Surge Withstand Capability
ANSI/IEEE C37.90. The F6000 Instrument functions as a source during
surge withstand capability tests, when the ANSI/IEEE-specified isolating
circuit is interposed between the instrument and the relay under test.
AC Amplitude Accuracy at 50/60 Hz
From 20
°
to 30
°
C: ±0.3% of setting or ±0.03% of range
From 0
°
to 50
°
C: ±0.5% of setting or ±0.05% of range
Typically 0.1% of reading
Distortion
Low distortion sine waves; total harmonic distortion 0.1% typical;
2% maximum at 50/60 Hz
Noise (10-30 kHz)
Voltage Source: 0.02% of range or 50 mV
Current Source: 0.02% of range or 1 mA