F6000 Family of Power System Simulators User Guide
72A-1589 Rev. C 02/01 G-7
D
DD
D
R
RR
R
A
AA
A
F
FF
F
T
TT
T
2
22
2
/
//
/
2
22
2
1
11
1
/
//
/
0
00
0
1
11
1
Logic Inputs
Eight galvanically isolated Logic Inputs, configurable as Voltage Sense or
Contact Sense.
Open Circuit Test Voltage: 12 VDC nominal
Short Circuit Test Current: 20 mA VDC nominal
Response Time: 0.1 millisecond maximum pick up and drop out
Isolation: ±500 V Peak
Triggers
Number: 8
Boolean combination of logic inputs can be used to define triggers.
Triggers are used to set timer start and stop conditions.
Timers
Number: 8
Accuracy: ±0.0005% of reading, ±50 microseconds
Resolution: 100 microseconds
Input Power Supply
115 V nominal at 15 A maximum (50 or 60 Hz)
230 V nominal at 10 A maximum (50 or 60 Hz)
Temperature
Operating temperature: 0
°
to 50
°
C
Storage temperature:
−
25
°
to +70
°
C
Humidity
Up to 95% relative humidity, non-condensing
Interfaces
RS-232 or Ethernet remote control to computer