PRELIMINARY
8.) Place the Power cable (the thinner, black cable) and the DVI cable (the thicker, grey cable) together
with the attached ferrite into their positions on the Monitor Arm assembly and secure them with the
white bracket, using the two Phillips screws previously removed,
as shown in Figure 8-128.
9.) Return the Keyboard Assembly to the system, as described in “Keyboard Assembly Installation
Procedure” on page 8-42.
10.)Thread the Power cable (the thinner, black cable) through the opening behind the axis of the
LCD arm and slide it upwards, see Figure 8-129.
11.)Thread the DVI cable (the thicker, grey cable) - together with the attached ferrite - through the
Figure 8-128 Bracket Securing Cables in Monitor Arm
Figure 8-129 Cables Threaded Past Monitor Arm Axis towards LCD