GE
P
ART NUMBER FN091065, REVISION 2 VS5 N AND VS6 N SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-151
PRELIMINARY
8-6-8 PSB (Probe Select Board) Board Replacement Procedure (cont’d)
7.) Loosen and remove the screw that secures the PIB-to-PSB flex cable in position and disconnect
the flex cable (see Figure 8-186).
8.) Squeeze together the two hinged latches (that secure the PSB to the cage assembly) to release the
spring locking mechanism - then remove the PSB Board (Figure 8-187).
Figure 8-186 Disconnecting the PIB-to-PSB Flex Cable
Figure 8-187 PSB Board Removed
PSB Board
Hinged Latch
Hinged Latch