IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
3-20 Section 3-5 - Completing the Setup
PRELIMINARY
The current drain will vary depending on the mains voltage.
At 230 VAC the current may be up to 2.2 A.
At 115 VAC the current may be up to 4.3 A.
At 110 VAC the current may be up to 4.6 A.
3-5-4 Connections on the I/O Rear Panel
NOTE: Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal
output part of Vivid™ S60/Vivid™ S70, configures a medical system, and is therefore responsible that
the Ultrasound system complies with the requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local representative for GE.
3-5-4-1 Connect Ethernet
Connect the network cable to the Ethernet connector on the External I/O.
The connector is located on the rear side of Vivid™ S60/Vivid™ S70.
3-5-4-2 Connect USB Flash Card
NOTE: USB Flash Cards approved for Vivid™ S60/Vivid™ S70 are verified for EMC performance according to
EN55011 class B. The use of any other USB Flash Cards will compromise this verification, and may
cause interference on Vivid™ S60/Vivid™ S70 itself, or on other electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Card in one of the USB ports on the Vivid™ S60/Vivid™ S70.
3-5-5 Connections on the Patient I/O Panel
The Patient I/O panel is located on the front of Vivid™ S60/Vivid™ S70.
Table 3-12 Electrical Specifications for all Vivid™ S60/Vivid™ S70 Models
Voltage 100-240 VAC ±10%
Power Consumption 500 VA
Frequency 50-60 Hz