IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
3-20 Section 3-5 - Completing the Setup
PRELIMINARY
The current drain will vary depending on the mains voltage.
• At 230 VAC the current may be up to 2.2 A.
• At 115 VAC the current may be up to 4.3 A.
• At 110 VAC the current may be up to 4.6 A.
3-5-4 Connections on the I/O Rear Panel
NOTE: Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal
output part of Vivid™ S60/Vivid™ S70, configures a medical system, and is therefore responsible that
the Ultrasound system complies with the requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local representative for GE.
3-5-4-1 Connect Ethernet
Connect the network cable to the Ethernet connector on the External I/O.
The connector is located on the rear side of Vivid™ S60/Vivid™ S70.
3-5-4-2 Connect USB Flash Card
NOTE: USB Flash Cards approved for Vivid™ S60/Vivid™ S70 are verified for EMC performance according to
EN55011 class B. The use of any other USB Flash Cards will compromise this verification, and may
cause interference on Vivid™ S60/Vivid™ S70 itself, or on other electronic devices.
For approved models, please refer to Chapter 9.
Insert the USB Flash Card in one of the USB ports on the Vivid™ S60/Vivid™ S70.
3-5-5 Connections on the Patient I/O Panel
The Patient I/O panel is located on the front of Vivid™ S60/Vivid™ S70.
Table 3-12 Electrical Specifications for all Vivid™ S60/Vivid™ S70 Models
Voltage 100-240 VAC ±10%
Power Consumption 500 VA
Frequency 50-60 Hz