IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
2-4 Section 2-2 - Console Requirements
PRELIMINARY
2-2-5-5 Power Stability Requirements
Voltage drop-out
Max 10 msec
Power Transients
The Vivid™ S60N/Vivid™ S70N ultrasound scanner is fully compliant with the following standard:
EN/IEC 60501-1-2 sections 36.202.1j, 7a(2).
2-2-6 EMI Limitations
Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies,
magnetic fields, and transients in the air or wiring. They also generate EMI. The Vivid S60N/Vivid S70N
ultrasound unit complies with limits as stated on the EMC label. However, there is no guarantee that
interference will not occur in a particular setup.
Note: Possible EMI sources should be identified before the unit is installed, and should not be on the
same line as the ultrasound system. A dedicated line should be used for the ultrasound system.
Electrical and electronic equipment may produce EMI unintentionally as the result of a defect. Sources
of EMI include the following:
Medical lasers.
Scanners.
Cauterizing guns.
•Computers.
•Monitors.
•Fans.
Gel warmers.
Microwave ovens.
Portable phones.
Broadcast stations and mobile broadcasting machines.

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals