IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-127
PRELIMINARY
8-5-16 TRx32 Boards Replacement Procedure
8-5-16-1 Tools
Appropriate Phillips screwdriver.
8-5-16-2 Time Required
15 minutes
8-5-16-3 Preparations
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-5-16-4 TRX32 Boards Removal Procedure
1) Remove the following covers: left side, right side, DVD, front.
2) Open and remove the Front End door assembly and place it on a stable surface:
3) Loosen the 2 screws that secure the metal cover of the TRX32 Box, and open the cover.
(Figure 8-153 below).
CAUTION
When performing these procedures, take precautions to avoid damage of
electrostatic-sensitive components. Always have the ESD wrist strap
connected either to the DIB chassis or to the GND plug at the rear of the
scanner, and to your hand.
If a battery is present, first remove the battery as it contains stored energy.
Refer to Battery Removal Procedure on page 8 - 145.
FRU Part #
Refer to Table 9-15 on page 9-14.
Air Inlet (Left Side) Cover Removal Procedure on page 8 - 7
Right Side Cover Removal Procedure on page 8 - 10
DVD Cover Removal Procedure on page 8 - 11
Front Cover Removal Procedure on page 8 - 16
Front End Door Assembly Opening and Removal Procedure on page 8 - 83
Figure 8-153 Opening the Metal Cover of the TRX32 Box