IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL 
8-34 Section 8-3 - Control Console Components - Replacement Procedures
PRELIMINARY
8-3-2 Touch Screen Service Cover Replacement Procedure
8-3-2-1 Tools
None.
8-3-2-2 Time Required
5 min
8-3-2-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-3-2-4 Touch Screen Service Cover Removal Procedure
Note: The Touch Screen Service Cover, located at the lower rear of the Touch Screen, clips into 
the Touch Screen Rear Cover by way of 4 securing tabs - Figure 8-34. 
1) Raise the console to the maximum up position. 
Make sure it is aligned in the central position (not pulled to one side or the other).
2) Using a flat screwdriver, carefully release the Service Cover snap-lock securing tabs (Figure 8-34) 
from the rear of the Touch Screen rear cover.
3) Using two hands, lift the service cover slightly to release it from the 4 securing tabs, then pull it 
forward to remove.
Refer to Table 9-9 on page 9-8 (Touch Panel Back Service Cover)
Figure 8-34   Touch Screen Service Cover - Securing Tabs
Figure 8-35   Removing the Touch Screen Service Cover
Snap-lock Tab
Snap-lock Tab
Snap-lock Tab