IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-166 Section 8-6 - Mechanical Platform Components - Replacement Procedures
PRELIMINARY
8-6-11 Probe Shelf Replacement Procedure
8-6-11-1 Tools
Use the appropriate screwdriver as indicated in the Probe Shelf Replacement procedure.
8-6-11-2 Time Required
45 min
8-6-11-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-6-11-4 Probe Shelf Removal Procedure
1) Remove the Keyboard Assembly as described in the Operator Panel Keyboard Assembly Removal
Procedure on page 8 - 44.
2) Working from below, and while carefully supporting the Probe Shelf Holder, remove the 4 retaining
screws - two on either side - as seen in Figure 8-201.
3.) Remove the Probe Shelf.
8-6-11-5 Probe Shelf Installation Procedure
1.) Return the Probe Shelf to its location on the Keyboard Interface Assembly and secure it with the
4 retaining screws (previously removed).
2.) Install the Keyboard Assembly as described in the Operator Panel Keyboard Assembly Installation
Procedure on page 8 - 47.
3) Lower the console to the maximum down position.
4.) Turn ON power to the system.
Refer to Table 9-7 on page 9-6.
Figure 8-201 Probe Shelf - Viewed from Below indicating Retaining Screws
Retaining Screw
Retaining Screw