IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
3-52 Section 3-6 - Configuration
PRELIMINARY
3-9-4-1 Wizard for Wireless Communication Setup
Note: The new wizard is intended for all systems with software version 201.46.0 or higher.
Table 3-20 Wizard for Wireless Communication Setup 1 of 3
Step Description Illustration
1 Turn on the Ultrasound System.
2 Connect the approved Wi-Fi dongle to the Ultrasound System.
3 On the touch panel, press Utility > Config.
4 Enter the Operator name (A), Password (B) and click Log on (C).
5
On the bottom tool bar, click the Connectivity button and then select
the Tcpip tab.