D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-174 Section 8-6 - Mechanical Platform Components - Replacement Procedures
PRELIMINARY
8-7-1-1 Preparation for Software Installation Procedures
NOTE: Make sure that the required Software Option keys are available prior to commencing the installation
process. Do not use the Software Option keys from a previous system version.
At the site, perform the following steps before you start the upgrade procedure:
1.) Perform Disk Management for all database records and select the "Copy" option
(for instructions, refer to the Vivid™ S60/Vivid™ S70 User Manual on your Document Media).
NOTE: It is recommended that Disk Management be executed using the "move" option to a network
repository.
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Vivid™ S60/Vivid™ S70 User Manual on your Document Media).
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
- TCP/IP address
- Network printer (if present)
- Wireless network settings (if present)
- Trackball speed
- InSite ExC settings
4.) When done, continue to Table 8-4 on page 8-175 and follow the Windows Operating System and
Vivid™ S60/Vivid™ S70 Software Installation Procedures.
8-7-1-2 Operating System and/or Application Software Installation Procedure
NOTE: A minimum of 1 hour is required to install both the Windows operating system and the Vivid™ S60 or
Vivid™ S70 application software. This excludes preparation time, backup etc. and performing functional
checks following the installation procedures.
Throughout the installation processes, the Vivid™ S60/Vivid™ S70 system will be unavailable
for scanning.
Table 8-4 on page 8-175 provides step-by-step instructions for installation of Windows operating
system software and Vivid™ S60/Vivid™ S70 ultrasound scanner software application. However, these
procedures may be performed separately, as required.
Follow the installation instructions below, as appropriate:
Windows operating system and Vivid S60N/Vivid S70N s/w application:step 1
Windows operating system ONLY: step 1 through step 10
Vivid S60N/Vivid S70N software application ONLY: step 11.
NOTE: It is necessary to perform the recommended Functionality Checks after completion of the installation
procedures, when installing the following:
Windows operating system and Vivid S60N/Vivid S70N s/w application
Vivid S60N/Vivid S70N software application ONLY
CAUTION
THIS PROCESS FORMATS THE HARD DRIVE (WHEN SELECTION OF THIS OPTION IS
APPLICABLE) - MAKE SURE TO SAVE ALL THE REQUIRED PRESETS, SETTINGS AND
PATIENT DATA!
WARNINGWARNING
REMOVE ALL EXTERNAL DEVICES SUCH AS PRINTERS AND USB CONNECTIONS
BEFORE STARTING THE UPGRADE PROCEDURE.

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals