D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-174 Section 8-6 - Mechanical Platform Components - Replacement Procedures
PRELIMINARY
8-7-1-1 Preparation for Software Installation Procedures
NOTE: Make sure that the required Software Option keys are available prior to commencing the installation
process. Do not use the Software Option keys from a previous system version.
At the site, perform the following steps before you start the upgrade procedure:
1.) Perform Disk Management for all database records and select the "Copy" option
(for instructions, refer to the Vivid™ S60/Vivid™ S70 User Manual on your Document Media).
NOTE: It is recommended that Disk Management be executed using the "move" option to a network
repository.
2.) Perform a full back-up for Patient Archive and System Configuration
(for instructions, refer to the Vivid™ S60/Vivid™ S70 User Manual on your Document Media).
3.) Write down the following settings as it will be necessary to restore them at the end of the process:
- TCP/IP address
- Network printer (if present)
- Wireless network settings (if present)
- Trackball speed
- InSite ExC settings
4.) When done, continue to Table 8-4 on page 8-175 and follow the Windows Operating System and
Vivid™ S60/Vivid™ S70 Software Installation Procedures.
8-7-1-2 Operating System and/or Application Software Installation Procedure
NOTE: A minimum of 1 hour is required to install both the Windows operating system and the Vivid™ S60 or
Vivid™ S70 application software. This excludes preparation time, backup etc. and performing functional
checks following the installation procedures.
Throughout the installation processes, the Vivid™ S60/Vivid™ S70 system will be unavailable
for scanning.
Table 8-4 on page 8-175 provides step-by-step instructions for installation of Windows operating
system software and Vivid™ S60/Vivid™ S70 ultrasound scanner software application. However, these
procedures may be performed separately, as required.
Follow the installation instructions below, as appropriate:
• Windows operating system and Vivid S60N/Vivid S70N s/w application:step 1
• Windows operating system ONLY: step 1 through step 10
• Vivid S60N/Vivid S70N software application ONLY: step 11.
NOTE: It is necessary to perform the recommended Functionality Checks after completion of the installation
procedures, when installing the following:
• Windows operating system and Vivid S60N/Vivid S70N s/w application
• Vivid S60N/Vivid S70N software application ONLY
THIS PROCESS FORMATS THE HARD DRIVE (WHEN SELECTION OF THIS OPTION IS
APPLICABLE) - MAKE SURE TO SAVE ALL THE REQUIRED PRESETS, SETTINGS AND
PATIENT DATA!
REMOVE ALL EXTERNAL DEVICES SUCH AS PRINTERS AND USB CONNECTIONS
BEFORE STARTING THE UPGRADE PROCEDURE.