D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-175
PRELIMINARY
Table 8-4 Windows Operating System and Vivid™ S60/Vivid™ S70 Software Installation Procedures
Step Instruction Expected Result
NOTE: Images in this procedure are for reference only. There may be differences or variations, in accordance with different software versions.
1.
Plug the Software Installation Media into the USB port
on the front of the Vivid™ S60N/Vivid™ S70N
ultrasound scanner.
2.
Press and hold down the <ON/OFF> button for
3 seconds.
The system powers ON and boot-ups from the Software Installation Media.
Windows starts and the message "Windows is loading Files...." is displayed on
the Touch Screen.
The System Software Loading...message is displayed with a progress bar:
When complete, the following is displayed on the Touch Screen:
This shows for which product the software is applicable and the versions (operating
system/application) and date, prompting you to select the required option.
3.
Click Install System SW A warning dialog box opens:
4.
Click Yes to continue.
Note:
Yes = performs a full format of the Hard Disk
(complete deletion of operating system and all
patient data)
No = only the operating system (Windows partition)
will be deleted; user data will remain
Another warning dialog box opens: