IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 1 - Introduction 1-9
1-4-3 Mechanical Safety
WARNINGWARNING
WHILE THE SOFTWARE INSTALL PROCEDURE IS DESIGNED TO PRESERVE DATA,
YOU SHOULD SAVE ANY PATIENT DATA, IMAGES, SYSTEM SETUPS TO A DVD OR
HARDCOPY BEFORE DOING A SOFTWARE UPGRADE.
WARNINGWARNING
PRIOR TO ELEVATING SCANNER, VERIFY THAT THE MONITOR IS LOCKED IN ITS
LOWEST POSITION. VERIFY THAT THE FRONT BRAKE IS LOCKED AND THE
SCANNER IS UNABLE TO SWIVEL. VERIFY THAT THE REAR BRAKES ARE IN THE
LOCKED POSITION.
DANGER
WHENEVER THE UNIT IS TO BE MOVED ALONG ANY INCLINE, USE EXTREME
CAUTION. MAKE SURE THAT THE VIVID™ S60N/VIVID™ S70N SCANNER AND ALL
PERIPHERALS ARE SECURELY MOUNTED IN PLACE BEFORE ATTEMPTING TO
MOVE IT.
DANGER
ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN
EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND
PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR
DEFECTIVE PROBE. FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN
SERIOUS INJURY AND EQUIPMENT DAMAGE.
DANGER
NEVER USE A PROBE THAT HAS FALLEN TO THE FLOOR. EVEN IF IT LOOKS OK, IT
MAY BE DAMAGED.
CAUTION
The Vivid™ S60N/Vivid™ S70N ultrasound scanner weighs 75kg (165 lbs.) or more, depending
on
carry-on peripherals when ready for use.
Care must be used when moving the unit or replacing its parts. Failure to follow the precautions
listed could result in injury, uncontrolled motion and costly damage.
ALWAYS:
Use two people when moving on inclines or lifting more than 16 kg (35 lbs)
Use the handle to move the system
Be sure the pathway is clear
Use slow, careful motions
Do not let the system strike walls or door frames
When moving the system on inclines, make sure that the Vivid S60N/Vivid S70N scanner and
all peripherals are securely mounted in place before attempting to move the scanner.
The rear handle should only be used for pushing the system. Do NOT use it for pulling the
ultrasound scanner (in any direction) as this may cause the system to become unstable in the
event of colliding with obstacles
WARNINGWARNING
THE SYSTEM SHOULD NOT BE MOVED WITH THE OPERATING PANEL EXTENDED.
POSITION THE OPERATING PANEL IN ITS CENTERED AND LOCKED POSITION.
LOWER THE OPERATING PANEL AS MUCH AS POSSIBLE BEFORE MOVING THE
SYSTEM. See Figure 5-7 on page 5-10 illustrating system in Transportation Mode.