D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
1-8 Section 1-1 - Overview
PRELIMINARY
WARNINGWARNING
WHEN THE TOP CONSOLE IS IN ITS LOCKED POSITION, THE GAS SHOCK IS
COMPRESSED AND STORES MECHANICAL ENERGY. DURING NORMAL OPERATION
THE TOP CONSOLE, THE WEIGHT OF THE MONITOR AND THE MECHANICAL FORCE
OF THE GAS SHOCK ARE IN BALANCE. TAKE CARE IF/ WHEN YOU ACTIVATE THIS
GAS SHOCK. PERSONAL INJURY CAN OCCUR AFTER THE PANEL IS REMOVED AND
THE SHOCK PRESSURE IS RELEASED. TAKE CARE WHEN YOU REPAIR THE
ELEVATION ASSEMBLY.
CAUTION
USE PROTECTIVE GLASSES DURING DRILLING, FILING AND DURING ALL OTHER WORK
WHERE EYES NEED PROTECTION.
CAUTION
USE SAFETY SHOES WHEN DOING WORK WHERE THERE IS ANY CHANCE OF FOOT
DAMAGE.
CAUTION
USE PROTECTIVE GLOVES WHEN DRILLING AND CUTTING.

Table of Contents