IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
3-2 Section 3-2 - Setup Reminders
PRELIMINARY
Section 3-2
Setup Reminders
3-2-1 Average Setup Time
The Vivid S60N/Vivid S70N setup and functional checkout will take approximately one hour; Vivid
S60N/Vivid S70N consoles with optional equipment (such as Alphanumeric Keyboard) may take slightly
longer.
Once the site has been prepared, the average installation time required is shown in Table 3-6 below.
3-2-2 Setup Warnings
1.) The Vivid S60N/Vivid S70N ultrasound scanner weighs 75 Kgs (165 lbs), without add-ons/
peripherals.
Two people are always required to unpack the system.
2.) There are no operator-serviceable components. To prevent shock, do not remove any covers or
panels. If problems or malfunctions occur, unplug the power cord. Only qualified service personnel
should carry out servicing and troubleshooting.
NOTE: For information on shipping carton labels, refer to Figure 3-7 on page 3-6 and Figure 3-10 on
page 3-8.
Table 3-6 Average Setup Time
Description
Average
Setup Time
Comments
Unpacking the scanner
30 minutes
Installing the scanner
30 minutes Time may vary, according to the required configuration
DICOM Option
(connectivity)
30 minutes Time may vary, according to the required configuration
Installing InSite
30 minutes

Table of Contents

Other manuals for GE Vivid S60N

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Vivid S60N and is the answer not in the manual?

GE Vivid S60N Specifications

General IconGeneral
BrandGE
ModelVivid S60N
CategoryMedical Equipment
LanguageEnglish

Related product manuals