IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-31
PRELIMINARY
Section 8-3
Control Console Components - Replacement Procedures
8-3-1 21.5" Monitor Replacement Procedure
8-3-1-1 Tools
Phillips screwdriver
8-3-1-2 Time Required
15 min
8-3-1-3 Preparation
1.) Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page
4 - 7.
1) Lower the console to the maximum down position.
Make sure it is aligned in the central position (not pulled to one side or the other).
2.) Fold the monitor forward into the face-down position, then lock the articulated arm.
8-3-1-4 Remove the 21.5 inch Monitor
1.) Remove the Monitor’s Rear Cover as described below:
a.) Remove the two fixing screws (Figure 8-32).
b.) Remove the Monitor’s Rear Cover.
FRU Part #
Refer to Table 9-9 on page 9-8.
Figure 8-32 Monitor Rear Cover’s fixing screews