D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 3 - System Setup 3-43
PRELIMINARY
3-8-4 Changing the AE Title and/or Port Number (Port No.)
Figure 3-31 AE Title and Port No.
1) To change AE Title and/or Port No., edit the respective fields.
2) Select Save settings to store your changes.
3) Reboot Vivid™ S60/Vivid™ S70 to activate the settings, or continue with other Tcpip set-up tasks.
3-8-5 Set the Remote Archive’s Network Information
To be able to connect to a remote archive on a remote computer or server, you must configure Vivid™
S60/Vivid™ S70 to communicate with it.
The configuration is done in the Server Config setup area on the Tcpip screen.
Figure 3-32 Server Config
3-8-5-1 To Access the Tcpip Screen
Follow this procedure to access the Tcpip screen:
1) Log on as ADM.
2) Select Connectivity > Tcpip.
3-8-5-2 To Add a Server in the Server Config List
Follow this procedure to add a server in the list:
1) Select Add.
2) Add the server's name in the Server Name field.
3) Add the servers IP address in the IP-Address field.
4) Select Check to verify that the server is found.
This check is a network Ping. A symbol to the right for the Check button indicates the result: