IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 3 - System Setup 3-41
PRELIMINARY
Section 3-8
Connectivity Setup
NOTE: If connected to a stand-alone network (Peer-to-Peer network with a Vivid™ S60/Vivid™ S70 scanner,
an EchoPAC workstation and an optional network printer), you should use default delivery settings.
3-8-1 Introduction
To be able to use the network functions when connected to a hospital network, the Vivid™ S60/Vivid™
S70 must have a proper network address.
• Before you can set up the Vivid™ S60/Vivid™ S70, you need to collect some information.
• The Worksheet (see sample Connectivity Installation Worksheet on page 2 - 13) can be used for
gathering this information.
• Typical source for this information is the network administrator.
3-8-2 Compatibility
Vivid™ S60/Vivid™ S70 can communicate with:
• EchoPAC
• Image Vault
For networks with Image Vault 5 server running on Windows Server 2003 or Windows Server 2008:
If not already done, install the latest version of the Vivid Raw Data Module (RDCM) on the Image
Vault server.
For instructions, see: “Vivid Raw Data Module (RDCM) Installation Manual”, Direction Number
IV294001.
• Other units via DICOM
NOTE: Vivid™ S60/Vivid™ S70 can’t read files directly from EchoPAC MAC or from MAC formatted MO disks.
Please use an EchoPAC PC workstation with an MO drive and with MacDrive 5 or MacDrive 6 installed.
EchoPAC PC can also open EchoPAC MAC examinations stored on a server.