D
IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-16 Section 8-2 - Covers - Replacement Procedures
PRELIMINARY
8-2-8 Front Cover Replacement Procedures
NOTE: If the RS cable restrainer option is installed, when replacing the Front Cover, it is necessary to also fit
the RS cable restrainer assembly. Note that for systems without the RS cable restrainer, when removing
the Front Cover, the steps for removing the RS cable restrainer assembly are not applicable.
8-2-8-1 Tools
Phillips screwdriver
8-2-8-2 Time Required
10 min
8-2-8-3 Preparations
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-2-8-4 Front Cover Removal Procedure
Note: The Front Cover is secured to the base of the mechanical platform by a screw through a
securing tab on each side of the cover (left and right inner side).
1) Remove all probes and ECG cables currently connected to the system.
2) Remove the front Basket for probe cable management - (hold with both hands and pull upwards).
3.) Remove the following covers: left side, right side, DVD.
Refer to Table 9-8 on page 9-7. (Front Assy Cover)
Figure 8-10 Front Cover - Securing Tabs