IRECTION FR091521, REVISION 1VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 4 - General Procedures and Functional Checks 4-7
PRELIMINARY
4-2-3 Power Shut Down
When switching the unit OFF, the system performs an automatic shutdown sequence.
When you switch off the unit, the system performs an automatic shutdown sequence.
Figure 4-1. Exit Dialog Window
The SYSTEM - EXIT menu, used when switching off the unit, gives you these choices:
Logoff
Use this button to log off the current user.
The system remains ON and ready for a new user to log on.
If the Logoff button is greyed out, it indicates that no user is logged on to the unit at the moment.
Shutdown
Use this button to shut down the system. The entire system will shut down.
If the Shutdown button is greyed out, use the key-combination <Ctrl+Alt+R> to shut down the unit.
•Cancel
Use this button to exit from the System-Exit menu and return to the previous operation.