IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 10 - Care and Maintenance 10-7
PRELIMINARY
10-5-2 Functional Checks
NOTE: Refer also to Chapter 4 -General Procedures and Functional Checks, for additional details about the
functional checks described in this section.
The functional checks take approximately 60 minutes to perform. Refer to the Vivid™ S60/Vivid™ S70
User Manual whenever necessary.
10-5-2-1 System Checks
Table 10-5 System Functional Checks
Step Item (or Mode) Description
1
B-Mode
Verify basic B-Mode (2D) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
2
CF-Mode
Verify basic CF-Mode (Color Flow Mode) operation.
Check the basic Ultrasound system controls that affect this mode of operation.
3
Doppler Modes
Verify basic Doppler operation (PW and CW if available).
Check the basic Ultrasound system controls that affect this mode of operation.
4
M-Mode
Verify basic M-Mode operation.
Check the basic Ultrasound system controls that affect this mode of operation.
5
Probe Elements
Perform an Element Test on each probe to verify that all the probe elements and system
channels are functional.
6
Applicable Software Options
Verify the basic operation of all optional modes such as Contrast. Check the basic
Ultrasound system controls that affect each option’s operation.
8
Operator Panel test Perform the Operator Panel Test Procedure.
9
Keyboard Do the interactive keyboard test.
10
Touch Panel Verify basic Touch Panel display functions.
11
Monitor Verify basic monitor display functions.
12
Peripherals See: Peripheral/Option Checks on page 10 - 8.