IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-64 Section 8-4 - Cables - Replacement Procedures
PRELIMINARY
7.) Squeeze the connector clips, then feed the display monitor cable lead to the OPIO through the
upper articulated arm joint as shown in Figure 8-76
.
8-4-1-4-1 Removing the Lower Arm
During this procedure, when removing the lower arm, the OPIO platform should be supported with
one hand while the Articulated Arm is lowered onto the system’s rear handle.
When performing this task, place protective sponge between the handle and the Articulated Arm as
shown in Figure 8-77 below.
1.) Remove the OPI:
Figure 8-76 Squeeze the Connector Clips and Feed Cable Downwards
Figure 8-77 Protective Sponge Between Articulated Arm and System Handle
• Operator Panel Keyboard Assembly Removal Procedure on page 8 - 44