IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 8 - Replacement Procedures 8-155
PRELIMINARY
8-6-7 Swivel and Up-Down Handle Replacement Procedure
8-6-7-1 Tools
Flat and Phillips screwdrivers as needed.
8-6-7-2 Time Required
60 min
8-6-7-3 Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in Power Shut Down on page 4 - 7.
8-6-7-4 Swivel and Up-Down Handle Removal Procedure
1) Remove the Operator Panel as described in the Operator Panel Keyboard Assembly Removal
Procedure on page 8 - 44.
2) Remove the Probe Shelf as described in the Probe Shelf Removal Procedure on page 8 - 166.
3.) Unscrew the protective metal plate (shown in Figure 8-189) and rotate it upwards 90 degrees.
FRU Part #
Refer to Table 9-5 on page 9-3 (Swivel Mechanism).
Figure 8-189 Protective Metal Plate Rotated 90 Degrees
Swivel Cable
Up-down Cable