IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
8-172 Section 8-6 - Mechanical Platform Components - Replacement Procedures
PRELIMINARY
3) Lift the chassis sufficiently to allow the wheel to drop down out of the wheel securing socket.
4.) Remove the wheel.
8-6-13-5 Brake and Locking Wheels Installation Procedure
1) Carefully lift the chassis sufficiently to allow insertion of the replacement wheel shaft into the wheel
securing socket.
2) Push the wheel shaft all the way up into the socket, then gently lower the chassis to the ground.
3.) Return and fasten the Allen screw (previously removed - see Figure 8-209), making sure the screw
is tightened sufficiently to secure the wheel on the chassis. Do not over-tighten the screw!
4) Return the screw cap (previously removed) by snapping it back into position (refer to Figure 8-208)