IRECTION FR091521, REVISION 1 VIVID S60N/VIVID S70N BASIC SERVICE MANUAL
Chapter 10 - Care and Maintenance 10-11
PRELIMINARY
10-5-4-2 Air Filter Cleaning
NOTE: For convenience (or if the air filter is excessively dirty), replacement filters are available. Refer
to Table 9-8 on page 9-7 for the air filter replacement part number.
Lock the Ultrasound system’s wheels prior to removing/cleaning the air filter. This prevents the
system from moving unexpectedly.
Table 10-9 Air Filter Cleaning - Frequency Varies with Your Environment
Step Item Description
1
Remove Filter
Cover
Remove the left side cover from the system, as described in Air Inlet (Left Side) Cover
Removal Procedure on page 8 - 7.
Remove the air filter from the cover, as shown in Figure 8-4 on page 8-8.
2
Clean Filter
The filter can be cleaned in sprinkling water, or it can be dusted with a vacuum cleaner. If
the filter is metal, wash and/or vacuum. If the filter is fiber or plastic, vacuum or replace. Dry
the filter.
3
Install Filter
Install the clean (and dry) filter. Refer to the Air Filter Installation Procedure on page 8 - 9 for
air filter installation instructions.