LIST (MULTI STEP MODE)
107
Setting range hold
When RANGE HOLD is set to OFF, the instrument will use the
automatic range to measure to get the correct amount.
When RANGE HOLD is set to ON, use the correct part for the first
measurement. The instrument will be fixed with the range used for
the first measurement as the range position to be measured later,
which can speed up the measurement.
When RANGE HOLD is set to OFF, the automatic range will be
restored for measurement.
However, if the fixed range is incorrect, it will result in a wrong value.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and press the
LIST SET function button on the right
side of the LCD to enter the [LIST SET]
page, and then press the Setup button
to enter [LIST MODE SETUP] page.
2. Use arrow keys to move cursor and
select RANGE HOLD item on the
[LIST MODE SETUP] page.
3. Use option key on the right of the LCD screen to
select a desired item.
Available
parameter
OFF
Use
automatic
range to measure
ON
Use fixed range for measure